Metrohm 719 S Titrino User Manual
Page 66
3.1. General rules
719 S Titrino
62
E120
Overrange of the primary measured value (pH, U, Ipol, Upol or T
with MEAS T). The secondary measured value (temperature) may
be instable as well.
Exit: Correct error or &m $S.
E123
Missing EP for calculation.
Exit: The error message disappears on next startup or on
recalculation.
E128
No new mean.
Exit: The error message disappears on next startup or on
recalculation.
E129
No new common variable, old value remains.
Exit: The error message disappears on next startup or on
recalculation.
E130
Wrong sample. For SET with preset titration direction the first
measured value lies behind the endpoint.
Exit: The error message disappears on next startup.
E131
No EP set for SET.
Exit: The error message disappears on next startup.
E132
Silo empty and it has been started with open silo or empty silo has
been opened.
Exit: Send a silo entry.
E133
Silo full.
Exit: Send new command.
E134
No method. A method, which is required from the silo memory, does not
exist.
Exit: The error message disappears on next startup.
E135
Check temp.sensor in MEAS T.
Exit: Correct error or &m $S.
E136
Same buffer in CAL. Measured value of the second buffer differs less than 6
mV from the measured value of the first buffer.
Exit: Correct error or &m $S.
E137
XXX Bytes are missing so that the method, the silo line could not be stored.
Exit: Send new command.