Metrohm 719 S Titrino User Manual
Page 32
2.7. Parameters, key
719 S Titrino
28
>stop conditions
stop V: abs.
**titr.
stop V 99.99 ml
**titr.
factor 999999
**titr.
filling rate max. ml/min
Stop conditions for titration
If this is not "normal", i.e. after reaching the EP.
Type of stop volume (abs., rel., OFF)
"abs.":
absolute stop volume in mL.
"rel.":
relative stop volume to sample size.
"OFF":
stop volume switched off. Stop volume
is not monitored.
If "abs." is set:
Absolute stop volume (0...9999.99 mL)
If "rel." is set:
Factor for relative stop volume (0...±999999)
Calculated as:
Stop V (in mL) = factor ∗sample size
Filling rate after the titration
(0.01...150 mL/min, max.)
The maximum rate depends on the Exchange Unit:
Exchange Unit
max.
5 mL
15 mL/min
10 mL
30 mL/min
20 mL
60 mL/min
50 mL
150 mL/min
>statistics
status: OFF
mean n= 2
res.tab: original
delete n= 1
Statistics calculation
Status of statistics calculation (OFF, ON)
If the statistics calculation is switched off, the following
inquiries regarding the statistics do not appear.
Mean value calculation from n single results (2...20)
Result table (original, delete n, delete all)
"original":
The original table is used. Deleted
individual results are again incorporated in
the evaluation.
"delete n": Deletion of single results with the index n.
"delete all": The entire table is deleted.
Delete data from sample number n (1...20)
The deleted result is removed from the statistics
calculation.