Metrohm 719 S Titrino User Manual
Page 92
3.2. Remote control commands
719 S Titrino
88
3.2.2.66. Info.ActualInfo.Assembly.Counter.V read
only
Info.ActualInfo.Assembly.Counter.Clear $G
$Q sends the volume. With the function &Info.Assembly.Counter.Clear $G, the
volume counter is set to zero.
3.2.2.67. Info.ActualInfo.Assembly.Meas read
only
Info.ActualInfo.Assembly.PistonPos read
only
$Q sends the current measured value from the assembly.
Piston position:
0
=
initial position (filled)
10 000 =
end position (empty)
3.2.2.68. Info.ActualInfo.Titrator.CyclNo read
only
Info.ActualInfo.Titrator.V read
only
Info.ActualInfo.Titrator.Meas read
only
Info.ActualInfo.Titrator.dVdt read
only
Info.ActualInfo.Titrator.dMeasdt read
only
Info.ActualInfo.Titrator.dMeasdV read
only
$Q sends the current values in the following formats:
SET
MEAS
CAL
CyclNo 127 127 127
V(ml) 1.2345
- -
Meas:
pH
U, Ipol (mV)
Upol (uA)
T (°C)
3.6(mV)
-345.6
-12.5
-
3.345
-345.6
-12.5
25.0
3.345
-
-
-
dVdt (ul/s)
2.5142
-
-
dMeasdt
pH,U,Ipol mV/s
Upol (uA/s)
T (°C/s)
0.7957
0.7957
-
0.7957
0.7957
0.7957
0.7957
-
-
dMeasdV (mV/ul) 10.6326
-
-
NV: Not Valid. If in the signal drift is OFF in modes MEAS and CAL, the signal
drift is NV.
OV will be sent for "overrange".
A time frame can be set up from the cycle number and the cycle time (see
3.2.2.71). The cycle number is set to 0 at the start of a method and it is in-
cremented until the end of the method.
3.2.2.69. Info.ActualInfo.EP.Index read
only
Info.ActualInfo.EP.X read
only
Info.ActualInfo.EP.Y read
only
.EP.X"1.234" Volume coordinate of the EP
.EP.Y"5.34" Measured value coordinate of the EP
3.2.2.70. Info.ActualInfo.Display.L1
up to 24 ASCII characters