Metrohm 719 S Titrino User Manual
Page 41
2.7 Parameters, key
719 S Titrino
37
sample changer cal: OFF
activate pulse: OFF
Calibration with sample changer (ON, OFF)
In calibrations with a sample changer, there are no
hold points in the calibration sequence for inputs.
Calibration temperature and pH values of the buffers
(which are temperature dependent) must therefore be
entered in advance. The inputs in key are
valid.
Pulse output on the line "activate" (L6, pin 1) of the
remote socket (all, first, OFF)
See page 131.
>statistics
status: OFF
mean n= 2
res.tab: original
delete n= 1
Statistics calculation
Status of statistics calculation (OFF, ON)
If the statistics calculation is switched off, the following
inquiries regarding the statistics do not appear.
Mean value calculation from n single results (2...20)
Result table (original, delete n, delete all)
"original":
The original table is used. Deleted
individual results are again incorporated in
the evaluation.
"delete n": Deletion of single results with the index n.
"delete all": The entire table is deleted.
Delete data from sample number n (1...20)
The deleted result is removed from the statistics
calculation.