2.8. Result calculations
719 S Titrino
40
RS1 text RS1
RS1 decimal places 2
RS1 unit: %
Text for result output (up to 8 characters)
Text input see page 8.
Number of decimal places for result (0...5)
Selection of result unit (%, ppm, g/L, mg/mL, mol/L,
mmol/L, g, mg, mL, mg/pc, s, mL/min, no unit or up to
6 characters).
Enter next formula, e.g. for RS2.
Meaning of the calculation variables CXX:
C00
Sample size, see page 50.
C01...C19
Method specific operands, see below. They are stored with the method
in the method memory.
C21...C23
Sample specific operands, see page 50ff.
C30...C39 Common
variables.
C40
Initial measured value of the sample, last measured value in MEAS.
C41 End
volume.
C42 Determination
time.
C43
Volume drift for SET with conditioning.
C44 Temperature.
C45
Dispensed start volume.
C46 Asymmetry-pH
(calibration).
C47
Electrode slope (calibration).
Input method specific operands C01...C19, key
With
in. For the calculation, the operands are used, which
were introduced in the formula.
The inputs method specific and are store in method
memory.
The calculation report can be printed with the key sequence:
C-FMLA
C
1