Metrohm 719 S Titrino User Manual
Page 81
3.2 Remote control commands
719 S Titrino
77
Assignment of common variables.
The values of the common variables are to be found in &Config.ComVar. They
can be viewed and entered there, see 3.2.2.50.
3.2.2.7. Mode.Def.Report.Assign
SET, MEAS, CAL:
full, short, calc, param, calib, ff
Definition of the report sequence, which is outputted automatically at the end
of the determination. Entries of more than one block have to be separated with
";".
3.2.2.8. Mode.Def.Mean.1.Assign RS1, RSX, EPX, CXX
Mode.Def.Mean.2.Assign
etc., up to .9
Assignment of the statistics calculations. Valid assignments are a requirement
for statistics calculations. In addition, the statistics calculation must be
switched on, see 3.2.2.22. Rules for statistics calculations see page 41.
3.2.2.9. Mode.CFmla
Mode.CFmla.1.Value
0...±999 999
Mode.CFmla.2.Value
etc., up to .19
Calculation constants specific to a method. Stored in the method memory of
the Titrino. Operands specific to the sample (3.2.2.52 and 3.2.2.53) and val-
ues of common variables (3.2.2.50) on the other hand are not stored with the
methods.
3.2.2.10. Mode.Parameter.SET1.EP pH:
0...±20.00, OFF
U,
Ipol:
0...±2000, OFF
Upol:
0...±200.0, OFF
Mode.Parameter.SET1.UnitEp read
only
Parameters for SET: Setting the 1st endpoint as pH value, in mV (with U and
Ipol) resp. µA (with Upol). The corresponding unit can be read with .UnitEP.
If the value is on "OFF", no further nodes will appear from SET1.
3.2.2.11. Mode.Parameter.SET1.Dyn pH:
0.01...20.00, OFF
U,
Ipol:
1...2000, OFF
Upol:
0.1...200.0, OFF
Mode.Parameter.SET1.UnitDyn read
only
Mode.Parameter.SET1.MaxRate 0.01...10...150, max.
Mode.Parameter.SET1.MinRate 0.01...25.0...9999.9
Parameters for SET: Control parameters, see page 25.
.Dyn:
Dynamics, control range in pH, mV (with U and Ipol) or µA (with
Upol). The corresponding unit can be read with .UnitDyn.
.MaxRate: Maximum allowed titration rate in mL/min. Max. means maximum
possible rate with the Exchange Unit in current use.
.MinRate: Minimum titration rate in ul/min.