4.2.
Diagnosis
719 S Titrino
111
Input 2
XX mV
• Carry out the same measurements as with Input 1.
• Press
analog input test 1...5
• Short-circuit input ‘Ind I’ (e.g. with cable 3.496.5070).
• Press <3>.
Input 1-2
XX mV
The differencial voltage between inputs ‘Ind I‘ and ‘Ind II’ is displayed.
Example: 0 - (+)1500 mV = -1500 mV
• Remove cables from the inputs ‘Ind I’ and ‘Ind II’.
• Press
analog input test 1...5
7.2 To check Pt 100 / 1000 connection
• Connect a Pt 100 or Pt 1000 sensor, a resistor switch-box or a single resistor of 100
Ω or 1 kΩ, respectively, to sockets ‘Pt 100/1000’ by means of short cables.
• Press <4>.
Pt 100*
XX °C
(* or Pt 1000)
Tolerance: ± 0.5 °C (Note also tolerance of resistor switch-box).
The sensor allows automatic displaying of the room temperature. (The resistor values corre-
spond to 0°C.)
• Press
analog input test 1...5
• Remove cables and resistor switch-box.
7.3 Polarizer
test
• Press <5>.
polarizer test
• Press
- 915 KF Ti-Touch (382 pages)
- 800 Dosino (53 pages)
- 767 Calibrated Reference (23 pages)
- 940 Professional IC Vario ONE/SeS/Prep 2 (54 pages)
- 754 Dialysis Unit (49 pages)
- 815 Robotic Soliprep for LC (76 pages)
- Vision Manual (207 pages)
- tiamo 2.1 Manual (1532 pages)
- 825 Lab Link (37 pages)
- 808 Titrando (70 pages)
- 902 Titrando (52 pages)
- 756 KF Coulometer (163 pages)
- 756 KF Coulometer (162 pages)
- 940 Professional IC Vario ONE/LPG (98 pages)
- 850 Professional IC Anion MCS Prep 3 (154 pages)
- 850 Professional IC Anion MCS Prep 3 (152 pages)
- 904 Titrando (58 pages)
- 850 Professional IC Anion MSM-HC MCS Prep 2 (150 pages)
- 930 Compact IC Flex Oven/ChS/Deg (47 pages)
- 872 Extension Module Liquid handling (64 pages)
- 814 USB Sample Processor (90 pages)
- 814 USB Sample Processor (91 pages)
- 940 Professional IC Vario (43 pages)
- Vision – Tutorial (40 pages)
- 799 GPT Titrino (242 pages)
- 889 IC Sample Center (68 pages)
- 761 Compact IC (228 pages)
- 851 Titrando (100 pages)
- 748 DH Sample Changer (32 pages)
- 940 Professional IC Vario ONE/SeS/HPG (51 pages)
- 896 Professional Detector – Amperometry (62 pages)
- 877 Titrino plus (139 pages)
- 881 Compact IC pro – Anion (129 pages)
- 940 Professional IC Vario ONE/ChS/HPG (112 pages)
- 930 Compact IC Flex Deg (41 pages)
- 840 PC Control 5.0 / Touch Control (351 pages)
- 940 Professional IC Vario ONE/Prep 1 (45 pages)
- 776 Dosimat (42 pages)
- 717 Sample Changer (36 pages)
- 815 Robotic USB Sample Processor XL (113 pages)
- 815 Robotic USB Sample Processor XL (114 pages)
- 940 Professional IC Vario ONE/SeS/PP (126 pages)
- 838 Advanced Sample Processor Installation Instructions (109 pages)
- 700 Dosino (55 pages)