719 S Titrino
104
silo empty
The silo memory is switched in but empty and a titration has
been started. Corrective action: At least the first 1 silo line
before starting the first titration.
Exit:
silo full
The silo memory is full up. Corrective action: If you have filled
less than 99 silo lines, you can create more space by deleting
old methods no longer needed. 1 silo line needs 18...90 bytes.
Exit:
stop time reached
SET has been stopped as the stop time has been reached.
stop V reached
The determination has been stopped as the stop volume has
been reached.
system error 3
The instrument adjustment data have been overwritten.
Exit:
message appears each time the instrument is switched on
until it has been readjusted (Metrohm service).
wrong sample
With SET, with preset titration direction the first measured
value is outside the end point.
Error messages in connection with the data transfer
If neither a computer nor a printer is attached, the report output at the end of the titration
must be switched off:
Receive errors:
E36
Parity
Exit:
same on both instruments
E37
Stop
bit
Exit:
same on both instruments
E38
Overrun error. At least 1 character could not be read.
Exit:
E39
Overflow of the receive buffer of the Titrino (> 82
characters).
Exit:
Send errors:
E40
DSR=OFF
E41
DCD=OFF
E42
CTS=OFF
Handshake unsatisfactory for more than 1 s.