4.5 Methods

■■■■■■■■■■■■■■■■■■■■■■

70

■■■■■■■■

743 Rancimat

Formula 3 for stability time:

on | off (Default value: off)

If this option is activated, the automatically determined stability time will
be calculated with the entered values. The result will be saved under the
name entered in the field "R3".

4.5.2.5

"Standards" tab

This function can be used to automatically calculate the induction time for
a temperature deviating from the measuring temperature, the so-called
standard time. The settings here do not have any effect on the measuring
sequence or the evaluation. After finishing the measurement, a post-cal-
culation or recalculation is possible at any time in the determination over-
view (see Chapter 4.7.5, page 121).

Calculate standard time

on | off (Default value: off)

Activation/Deactivation of the calculation of the standard time according
to the following formula:

Standard time = Factor

(Temperature - Target temperature)/10

x induction time

Formula according to...

Selection

AOCS Cd12b-92 | AOM | free