3 graph and reevaluation, Graph and reevaluation – Metrohm 743 Rancimat User Manual
Page 120
4.7 Results
■■■■■■■■■■■■■■■■■■■■■■
112
■■■■■■■■
743 Rancimat
Directory
Directory in which the report has been saved.
Description
The following warning messages are possible for incorrectly executed
determinations:
Temperature not reached: setpoint temperature at start of the deter-
mination not reached.
Missing measuring points (start time): measuring points are missing
from the given time (in s) onwards.
Missing measuring points (end time): measuring points are present
again from the given time (in s) onwards.
Manual Stop: the determination was stopped manually with [Stop].
Stop after communication error: Automatic stop of the determination
because of interrupted connection between the Rancimat and the PC.
Value
Value for warning (e.g. temperature, time).
4.7.3
Graph and reevaluation
You can either open a graph of an individual determination (single
graph) or have graphs of several determinations displayed above one
another (multiple graph). In addition, you can open the so-called live
graph of a determination, in the dialog window of which you can man-
ually perform reevaluations of measuring curves.
You have the following options for reevaluation of graphs:
■
Setting the induction time manually
■
Setting the stability time manually
■
Setting tangents manually