beautypg.com

4 carrying out glp tests, 1 temperature, Carrying out glp tests – Metrohm 743 Rancimat User Manual

Page 153: Figure 7, Accessories for glp test temperature

background image

■■■■■■■■■■■■■■■■■■■■■■

4 Operation

743 Rancimat

■■■■■■■■

145

4.8.4

Carrying out GLP tests

4.8.4.1

Temperature

For the GLP test for temperature measurement the calibrated tem-
perature sensor and the measuring insert are required (part of the option-
ally available GLP test set, see Optional accessories, page 185). For this,
the temperature sensor must be connected to the Pt100 connector on the
rear of the 743 Rancimat.

2

1

5

4

3

Figure 7

Accessories for GLP test Temperature

1

Pt100 temperature sensor

2

PTFE cylinder

3

Hexagon screw

4

Hexagon key (6.2621.120)

5

Aluminum cylinder

Preparing the GLP test set

Proceed as follows in order to prepare the GLP test set according to Figure
7
:

1 Assemble accessories

Insert the temperature sensor from above into the opening of the
PTFE cylinder.

Place the PTFE cylinder on the aluminum cylinder in such a way
that the temperature sensor fits into this opening.

Push the temperature sensor into the aluminum cylinder as far as
the limit stop.