4 calibration functions, 1 determining cell constants, Calibration functions – Metrohm 743 Rancimat User Manual

Page 56: Determining cell constants

background image

4.4 Calibration functions

■■■■■■■■■■■■■■■■■■■■■■

48

■■■■■■■■

743 Rancimat

Time

Time of the event. The format depends on the settings defined in Win-
dows in Control Panel

Country settings Time.

User

Name of the user logged in at the time of the event.

Unit

Instrument number (1…4).

Event

Description of the event.

4.4

Calibration functions

4.4.1

Determining cell constants

Only the change in conductivity is measured an evaluated during the
Rancimat measurement. As a result, the cell constants of the conductivity
measuring cells used in the Rancimat, which are normally in the range
from 1.00...1.20, only have to be entered exactly if the absolute con-
ductivity measured value
is to be displayed correctly.

The value also has to be entered before carrying out the GLP test "Con-
ductivity"
. The cell constants can thereby be entered manually or deter-
mined automatically using a defined standard solution (see Optional
accessories, page 185)
.

Entering cell constants manually

Proceed as follows if you wish to enter the cell constants manually:

1 Open dialog window

Click on the menu item Tools

Cell constants Manual....

The following dialog window appears: