6 Troubleshooting

743 Rancimat

■■■■■■■■

167

Problem

Cause

Remedy

The induction time
is longer/shorter
than expected.

The temperature is not
selected correctly.

Ensure that the correct method for the
determination has been selected.

Check that Temperature and Delta T
have been specified correctly in the
method (see Chapter 4.5.2.1, page 60).

See also: The induction
times are not reproducible
for multiple determina-
tions.

The stability time is
longer/shorter than
expected.

Delta Kappa is not defined
correctly.

Ensure that the value of Delta Kappa defined
in the method is correct (see Chapter 4.5.2.2,
page 64)
.

See also: The stability times
are not reproducible for
multiple determinations.

See also: The induction
time is longer/shorter than
expected.

The measurement
curves are extremely
noisy.

The air supply for the
measuring solution is direc-
ted to the conductivity
measuring cell.

Loosen tubing adapter (5-

11

) on the measur-

ing vessel cover, turn the PTFE tube (5-

17

) in

such a way that the air is no longer directed to
the measuring cell and fix in this position.

Gas bubbles adhere to the
conductivity measuring cell
during measurement.

Ensure that the measuring cell is clean and
free of fat. If necessary, clean thoroughly
(see Chapter 4.6.5, page 83).

In some cases ultra pure water contains a
large proportion of dissolved air which
escapes during measurement. In this case,
degas the ultra pure water before the
measurement for 5 to 10 min in a vacuum.

During the measurement,
sample evaporates in the
reaction vessel and con-
denses in the measuring
vessel. This results in con-
tamination of the conduc-
tivity measuring cell, which
in turn promotes adher-
ence of gas bubbles.

Keep the measuring time as short as possi-
ble, by about 4 to 6 h. The induction time
can be reduced by about half by increasing
the temperature by 10 °C.