beautypg.com

Metrohm 743 Rancimat User Manual

Page 61

background image

■■■■■■■■■■■■■■■■■■■■■■

4 Operation

743 Rancimat

■■■■■■■■

53

Click on the symbol .
The dialog window with the method parameters is opened.

Under Temperature enter the required setpoint temperature for
the sample and set Delta T to 0 °C.

Set the gas flow to the required value and finalize with [Save].

5 Start gas flow and heating

Click on the menu item Tools

Gas flow control Gas flow

on.

Click on the button [Heat to xx °C].

Executing a Delta T determination

After you have prepared everything for determining Delta T (see "Prepar-
ing Delta T determination", page 52)
, you can now execute the determi-
nation of Delta T as follows:

1 Enter sensor calibration data

Click on the menu item Tools

Determine Delta T Block A

or Block B.
The following dialog window appears:

Enter a Sensor name.

Enter the calibration data (temperature and resistance) for Points
1