4.8 GLP functions
■■■■■■■■■■■■■■■■■■■■■■
144
■■■■■■■■
743 Rancimat
Tolerance
Indication of the tolerance value for the relevant GLP test. If the result is
within the range Setpoint ± Tolerance, the test is regarded as passed. If
you click on [Set default], the specified standard value is entered in this
field.
Additional alert message text
Additional message text which is to be added during the prompt for carry-
ing out the next GLP test.
4.8.3
GLP status
Displaying the GLP status
Proceed as follows to display the status of the GLP monitoring:
1 Open dialog window
■
If several instruments are connected: In the Control window,
use the menu item View
▶ Unit x to select the required instru-
ment.
■
Click on the menu item Tools
▶ GLP test ▶ Status....
The dialog window GLP status for unit with serial no. x appears.
Parameter description
Test type
Type of GLP test (Temperature, Conductivity or Gas flow).
Next test due date
Date on which the next GLP test is to be carried out. If the GLP monitoring
is not activated, Disabled appears in this column.
Due in
Number of days until the next GLP test.
Last test result
Indication whether the test has been passed (Ok) or not (Not Ok!).