4.8 GLP functions

■■■■■■■■■■■■■■■■■■■■■■

150

■■■■■■■■

743 Rancimat

4.8.4.2

Conductivity

The GLP test for conductivity measurement can be carried out either
with a standard solution with a known conductivity or with the test
resistor 6.2109.030
. The test resistor is part of the optionally avaiable
GLP test set (see Optional accessories, page 185). The test resistor must
be connected during measurement at the electrode connector (3-2) of the
743 Rancimat.

Carrying out the GLP test for the conductivity measure-
ment

Proceed as follows to carry out the GLP test:

Note

Before carrying out the GLP test with standard solution for the first
time, the cell constants of the conductivity measuring cells must be
determined (see Chapter 4.4.1, page 48).

1 Select instrument

In the Control window, use the menu item View

Unit x to

select the instrument for which the GLP test is to be carried out.

Carrying out the GLP test with the test resistor: continue with
step 2.
Carrying out the GLP test with the standard solution: continue
with step 3.