beautypg.com

7 technical specifications, 1 general data, 2 temperature regulation and measurement – Metrohm 743 Rancimat User Manual

Page 178: General data, Temperature regulation and measurement

background image

7.1 General data

■■■■■■■■■■■■■■■■■■■■■■

170

■■■■■■■■

743 Rancimat

7 Technical specifications

7.1

General data

Number of instru-
ments that can be
connected

1…4

Number of sam-
ples per instru-
ment

1…8 (4 each per heating block)

Sample amount

Several grams / several milliliters

7.2

Temperature regulation and measurement

Range setpoint
temperature

50…220 °C

Temperature cor-
rection (Delta T)

0…±9.9 °C
(can be entered manually or determined automatically using the exter-
nal temperature sensor)

Temperature
measuring range

0…250 °C

Resolution

0.1 °C

Max. deviation of
the heating block
temperature from
the set value
(50…220 °C)

±0.3 °C

Reproducibility of
the set tempera-
ture

typ. < 0.2 °C

Max. temperature
difference
between different
measuring posi-
tions per block

typ. < 0.3 °C

Temperature fluc-
tuations

typ. < 0.1 °C
(at attained operating temperature, inserted and identically filled reac-
tion vessels and 20 L/h air throughput)