4 Operation

743 Rancimat

■■■■■■■■

65

Evaluate induction time

on | off (Default value: on)

Determination of the induction time (see "Induction time", page 66).

Evaluate stability time

on | off (Default value: off)

Determination of the stability time (see "Stability time", page 67).

Note

If you select this evaluation mode, you always need to calibrate the
conductivity measuring cell beforehand.

Delta Kappa

Conductivity change for the determination of the stability time.

Range

1 ... 400 µS/cm

Options for induction time

Evaluation delay

Waiting time until starting the curve evaluation.