background image

4.4 Calibration functions

■■■■■■■■■■■■■■■■■■■■■■

54

■■■■■■■■

743 Rancimat

Note

You can find the calibration data (T1 to T3) for Metrohm sensors
on the certificate enclosed with the sensor or on the label fastened
on the sensor cable.

The following dialog window appears:

2 Wait for measurement

Wait until the message Measured temperature is now stable!
appears in the message box.

The following three values are displayed below the message box:

Block temperature

Temperature which has been measured in the
heating block with the internal temperature
sensor.

Temperature of ext. sen-
sor

Temperature which has been measured in the
reaction vessel with the external temperature
sensor.

Delta T

Temperature correction which is determined
from the difference between the block temper-
ature and the temperature of the ext. sensor.