background image

4.8 GLP functions

■■■■■■■■■■■■■■■■■■■■■■

146

■■■■■■■■

743 Rancimat

2 Screw on accessories

Use the hexagon key to lightly tighten the hexagon screw.

Carrying out the GLP test for the temperature measure-
ment

Proceed as follows to carry out the GLP test:

Note

Before starting the GLP test, ensure that the temperature of the inser-
ted GLP test set is stable. If the GLP test set is used in a cold heating
block, this condition is fulfilled after the normal heating up time of the
heating block and the GLP test set. If the GLP test set, on the other
hand, is used in an already heated heating block, it is necessary to wait
at least 30 min before starting the GLP test.

1 Select instrument and method

In the Control window, use the menu item View

Unit x to

select the instrument for which the GLP test is to be carried out.

Under Unit x Block A and B, click on the symbol in order to
select the method whose temperature is required for the test.

If required, adjust the temperature (see "Opening a selected
method and adjusting parameters", page 57)
.