beautypg.com

Figure 4 – Metrohm 743 Rancimat User Manual

Page 19

background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

743 Rancimat

■■■■■■■■

11

Molecular sieve flask

Filter

Air out

2

Air/N in

From

flask

To
flask

8

9

10

12

11

1

2

3

6

5

4

7

Figure 4

Mounting accessories for the air supply (rear of the instru-
ment)

1

Drying flask cover (6.1602.145)
Cover for the drying flask.

2

Drying flask (6.1608.050)

3

Flask holder
For fastening the drying flask.

4

Filter tube (6.1821.040)

5

FEP tubing 250 mm (6.1805.080)
For supplying the air from the internal pump
to the drying flask.

6

FEP tubing 250 mm (6.1805.080)
For supplying the air from the drying flask to
the reaction vessel (2-

11

).

7

"From Flask" connection

8

Dust filter (6.2724.010)

9

"To Flask" connection

10 "Air out" connection

11 FEP tubing 130 mm (6.1805.010)

12 "Air/N

2

in" connection

Mount the accessories for the air supply as follows:

1 Mount dust filter

Insert the dust filter (4-8) on the connection marked with Filter
on the rear of the Rancimat.

If the laboratory air is heavily contaminated, a tubing for supply-
ing fresh air can be connected to the dust filter.