6.1 Problems
■■■■■■■■■■■■■■■■■■■■■■
164
■■■■■■■■
743 Rancimat
6 Troubleshooting
6.1
Problems
Problem
Cause
Remedy
The pump is louder
than normal.
The air flow is blocked
somewhere before or after
the pump.
■
) and, if necessary,
replace.
■
Check filter tube (4-
cover for blockages and, if necessary, free
with a gentle tap.
■
Examine FEP tubings (2-
and (3-
) for blockages and, if necessary,
replace.
Extra air is sucked in at
somewhere else other than
the dust filter. There is a
leak somewhere in the sys-
tem before the pump.
■
) for cracks, kinks etc. and tighten. If
necessary, replace.
■
Place the drying flask cover (3-
) correctly
on the drying flask and screw tight.
No air flow can be
detected in the reac-
tion vessel (it does
not bubble),
although the pump
is running.
The air supply is blocked.
■
Remove FEP tubing (5-
adapter (5-
). A slight air flow must be dis-
cerned here. If this is not the case, please
contact Metrohm Service.
■
) for blockages. If
necessary, clean or replace.
■
) and air tube
) on the reaction vessel cover for
blockages. If necessary, clean or replace.
The FEP tubing for the air
supply is defective.
Check FEP tubing (2-
necessary, replace.
The FEP tubing for the air
supply is not connected
correctly.
Tighten FEP tubing (2-
) on both sides.
The air tube (5-
) does not
immerse in the sample.
■
Push reaction vessel cover (5-
) down as
far as the limit stop.
■
Use more sample.