2 820 ic separation center, 820 ic separation center – Metrohm 820 IC Separation Center User Manual
Page 98
6 Appendix
819 IC Detector / 820 IC Separation Center
90
Height
130 mm
Depth
343 mm
Weight
8.2 kg (with detector block, without accessories)
6.1.2
820 IC Separation Center
Temperature control of the column heating
Temperature range
Ambient temperature +5°C...80°C, adjustable in
steps of 0.1 °C.
Accuracy
±
0.5 °C
Stability
better than 0.05 °C
Parts and controls
Indicators
Power LED 11 for Separation Center
Power LED 12 for Column heating
Power supply of separation center
Internal
Internal supply of 819 IC Detector via connecting
cable
External
External supply via DIN socket with 6.2152.000
power supply unit:
5 V / 0.5 A
24 V / 2 A (transient, 200
…
300 ms)
Power supply of column heating
Via jack bush with 6.2152.010 power supply unit:
24 Vdc / 2 A
Interfaces
Valve interfaces
Specifications, see Section 5.5
Safety specifications
Construction/testing
According to EN/IEC 61010-1 / UL 3101-1,
Protection class III
Safety notes
The Instructions for Use include information and
warnings which must be heeded by the user to
assure safe operation of the instrument.
Electromagnetic compatibility (EMC)
Emitted interference
Standards met:
- EN/IEC 61326
- EN 55022
- CISPR 22