5 interfaces – Metrohm 820 IC Separation Center User Manual
Page 92
5 Interfaces
819 IC Detector / 820 IC Separation Center
84
819 IC Detector
Pin
Function
Outputs
5 Valve
A
→
"FILL"
Valve A is switched to position "FILL".
H
L
t
P
> 50 ms
18 Valve
A
→
"INJECT"
Valve A is switched to position "INJECT".
H
L
t
P
> 50 ms
4
Lock Terminal block 26, 31
L = functions
H = functions
V
CEO
= 40 V
IC =
20
mA
17 Valve
B
→
"FILL" or suppressor
→
"STEP"
Valve B is switched to position "FILL" or suppressor
module is switched to next position.
H
L
t
P
> 50 ms
3 Valve
B
→
"INJECT"
Valve B is switched to position "INJECT".
H
L
t
P
> 50 ms
2 Remote
6
Programmable remote output line 6
(command "remote" in program or event)
V
CEO
= 40 V
IC =
20
mA
1 Remote
7
Programmable remote output line 7
(command "remote" in program or event)
V
CEO
= 40 V
IC =
20
mA
16 Remote
8
Programmable remote output line 8
(command "remote" in program or event)
V
CEO
= 40 V
IC =
20
mA
Voltages
+5V
0V
24VAKT
0VAKT
15
14
25
6
13
8
19
20
5 V: inactive
High (H)
0 V: active
Low (L)
I
≤
40
mA
R
i
≅
12
Ω
Contact arrangement at socket
"IC Separation Center" (female)
13
1
25 14
Connecting cables to external devices must have an appropriate
25-pin connector (male). Metrohm offers connector sockets
(K.210.9060) and the associated housing (K.210.0002) as an
option for the preparation of such cables.
No liability whatsoever will be accepted for damage arising from the improper connection of devices.