3.2 820 IC Separation Center
819 IC Detector / 820 IC Separation Center
65
Only active if the thermostat is off. Switch the
thermostat of the column heating on and apply
the operating temperature specified under
Set
point
, the button turns to
.
Only active if the thermostat is on and a new
operating temperature is entered at
Set point
.
Sets the new operating temperature.
Only active if the thermostat is on. Switch the
thermostat of the column heating off.
Start with startup hardware
Start the
Thermostat
with the startup of the sys-
tem hardware.
Enabled
The temperature of the column heating
is
monitored
by
«
IC Net
»
, an injection is trig-
gered, if the set operating temperature of the
column heating is reached.
The operating temperature is reached, when
the actual temperature is within
±
1 °C of the
temperature set above.
Disabled
The temperature of the column heating
is not
monitored
by
«
IC Net
»
. The proceeding of a
determination or a queue is not affected if the
set value is not reached.
Links
The
Links
tab of the
820 IC Separation Center
is used for COM port
selection and settings (details see Section 5.2.4 Links of the «IC
Net» Instructions for Use), resp. for Event line selection.