Fig. 29, Assembling the suppressor, Fig. 29 ) – Metrohm 820 IC Separation Center User Manual
Page 82: 72 from suppressor holder 75, Pull out connection piece 73 and suppress, R 74 from, R holder 75 (the connection, Ection piece 73 from suppress, R 74, N piece 73
4 Notes - Maintenance - Faults
819 IC Detector / 820 IC Separation Center
74
6 Clean connection piece
•
Clean the sealing surface of connection piece 73 using a lint-
free cloth and ethanol.
Fig. 29:
Assembling the suppressor
72 Screw nut
73 Connection piece with input and
output leads
(6.2832.010)
for suppressor module
74 Suppressor rotor(6.2832.000)
75 Suppressor holder
7 Insert connection piece
•
Insert connection piece 73 in suppressor holder 75 in such a
way that connection "1" is at the top, and that the three lugs
on the connection piece fit in the corresponding openings of
the holder.
•
Screw nut 72 onto the thread of suppressor holder 75 manu-
ally (do not use tools).
8 Connect and condition the suppressor
•
Reconnect the suppressor to the IC system.
•
Before switching the suppressor to the next position for the
first time, rinse all 3 suppressor units with solution for 5 min.
74
73
72 75