2 installation of accessories, 1 insert detector block, Installation of accessories – Metrohm 820 IC Separation Center User Manual
Page 23: Insert detector block, Fig. 7, Insert detector block in the separation center
2.2 Installation of accessories
819 IC Detector / 820 IC Separation Center
15
2.2 Installation
of
accessories
2.2.1
Insert detector block
In order to operate an 820 IC Separation Center with an 819 IC Detector
the detector block contained in the standard equipment of the detector
(1.732.0110 Detector block metal-free or 1.732.0100 Metal) must
be inserted in the upper part of the Separation Center.
Fig. 7:
Insert detector block in the Separation Center
2.820.0X10/2.820.0X30 IC Separation Center
The instrument versions 2.820.0X10 and 2.820.0X30 of the IC Separa-
tion Center are operated with one 819 IC Detector. It is best to proceed
as follows when connecting the two instruments and the detector block:
1 Install detector block
•
Unscrew the four knurled screws 15 from the top rear panel
21 of the 820 IC Separation Center and remove rear panel
•
Position detector block from the back in the space provided
in the 820 IC Separation Center on the right and push fully to
the front (see Fig. 7, resp. Fig. 21 and Fig. 23).
•
Insert the cable permanently attached to the detector block in
opening 20 and the outlet capillary in opening 19 "Waste A"
of the rear panel 21.
•
Replace rear panel 21 and screw to the 820 IC Separation
Center using the four knurled screws 15.