Metrohm 820 IC Separation Center User Manual
Page 40
2 Installation
819 IC Detector / 820 IC Separation Center
32
3 Installation of the capillary in the IC Separation
Center
•
Unscrew the four knurled screws 15 of the bottom rear panel
28 of the 820 IC Separation Center and remove rear panel
•
Insert PEEK capillary 48 from the back into the inner com-
partment of the IC Separation Center.
•
Install rear panel 28 so that the capillary is positioned in
opening 34 „Inlet A“ or 32 „Inlet B“ and screw on with the four
knurled screws 15.
4 Connection of the pulsation dampener
•
Procedure see section 2.7.2.
5 Connection to the injection valve
•
Connect PEEK capillary 48 to pulsation dampener 49 (see
Section 2.7.2). Using another PEEK capillary 48, connect pul-
sation dampener to connection „5“ (see Fig. 28) of injection
6 Mount column connection capillary
This step is only necessary, if no column heating is used.
•
Connect column connection capillary50 (ca. 20 cm of
6.1831.010 PEEK capillary) to connection „4“ (see Fig. 28) of
injection valve 51.