3 Operation
819 IC Detector / 820 IC Separation Center
62
Auto Step
*
fill, inject, off
Display of current auto step status.
Entry field
Set automatic switching of the suppressor
module after each "fill" or "inject" of valve A.
Selection:
fill, inject, off
* only available for
820 IC Separation Center
connected to 819
Fig. 28:
Flow path of the injection valve
If the injection valve is in the
position, the sample loop 61 is
filled by siphoning in the sample solution using the syringe. The eluent
flows directly to the separating column.
If the injection valve is in the
position the eluent flows via the
sample loop to the separating column. This causes injection of the
sample previously added to the sample loop. Sample inlet tubing and
syringe are directly connected.
1
2
4
5
3
6
Syringe
Column
Eluent
Sample
Loop
1
2
4
5
3
6
Syringe
Column
Eluent
Sample
Loop