2.4 Mains connection
819 IC Detector / 820 IC Separation Center
25
2.4 Mains
connection
Follow the instructions below for connecting to the power supply. If
the instrument is operated with a mains voltage set wrongly and/or
wrong mains fuse, there is a danger of fire!
2.4.1
819 IC Detector
Setting the mains voltage
Before switching on the 819 IC Detector for the first time, check whether
the mains voltage set on the instrument (see Fig. 15) matches the local
mains voltage. If this is not the case, you must reset the mains voltage
on the instrument as follows:
1 Disconnect mains cable
Disconnect mains cable from mains connection plug 3 of the
819 IC Detector.
2 Remove fuse holder
Using a screwdriver, loosen fuse holder 4 below the mains con-
nection plug 3 and take out completely.
3 Change and check fuse if necessary
Carefully take the fuse installed for the desired mains voltage out
of fuse holder 4 and check its specifications (the position of the
fuse in the fuse holder is marked by the white arrow imprinted
next to the mains voltage range):
100
…
120 V 0.63 A (slow blow) Metrohm-No.
U.600.0014
220
…
240 V 0.315 A (slow blow) Metrohm-No. U.600.0011
4 Insert fuse
Change fuse if necessary and reinsert in fuse holder 4.
5 Install fuse holder
Depending on the desired mains voltage, insert fuse holder 4 in
the 819 IC Detector so that the corresponding mains voltage
range can be read normally and the adjacent white arrow points
to the white bar imprinted below the fuse holder (see Fig. 15).