3.1 819 IC Detector
819 IC Detector / 820 IC Separation Center
51
Conductivity
Live display of measured conductivity (auto-
zero value) in
µ
S/cm. This field is available for
the
WATCH WINDOW
. The watch window color is
changed to the
Out of range
color if the meas-
ured value is more than 80% outside the
Full
Scale
range
.
abs. Cond.
Live display of absolute conductivity value in
µ
S/cm. This field is available for the
WATCH
WINDOW
. The watch window color is changed
to the
Out of range
color if the measured value is
bigger or equal to 110% of
Range.
Ion
Display of ion settings:
Anion
:
polarity = +, temperature coeff. = 2.5
Cation
:
polarity = –, temperature coeff. = 1.5
Range
Display of measuring range in
µ
S/cm.
Full
scale
Display of full scale range in
µ
S/cm.
Cell
const
. Display
of cell constant of the conductivity cell
in /cm.
Thermostat
Display of temperature within in the detector
block.
Offset
Display of offset of the zero point.
Damping
Display of electronic damping of the analog
output.
Remote lines
Display value of remote line settings 1
…
8.