6.1 Technical data
819 IC Detector / 820 IC Separation Center
87
6 Appendix
6.1 Technical
data
Provided that nothing to the contrary is mentioned, the
published data are typical values at an ambient tem-
perature of 25°C.
6.1.1
819 IC Detector
Conductivity measurement
Measurement ranges
100, 200, 500, 1000, 2000, 5000, 10'000
µ
S/cm
Full Scale
0.05
…
10'000
µ
S/cm
Temperature correction
Automatic correction of the measured conductiv-
ity to reference temperature 20°C with the adjust-
able temperature coefficients
(1.5 %/°C or 2.5 %/°C)
Accuracy, absolute value
< 4 % of full-scale value for all ranges under the
following reference conditions:
Adjustment with reference resistance to 1.000 V
(range 1 mS/cm, cell constant 16.7 /cm, temp.
coeff. 2.5 %/°C, frequency 5 kHz, room tempera-
ture) to accuracy <
±
2 mV, measurement accu-
racy < 0.5 % after 30 min warm-up time
Linearity
Up to
±
150 % of each full-scale range
Deviations < 0.5 % of full-scale range
Temperature dependence
Typically 25 ppm/°C
Dependence of cell const..
<
±
1 % over entire adjustment range
Drift (electronic)
< 0.0013 % of selected measurement range
/h/°C
Noise (electronic)
Typically < 0.0003 % of selected meas. range
Damping
2-stage damping (Bessel 4
th
order)
"
Damping off
"
0.25 s (10
…
90 %)
"
Damping on
"
2.00 s (10
…
90 %)
Autozero
Function
Automatic zero setting (electronic background
compensation) over entire selectable measure-
ment range
Maximum error
±
0.6 % of full-scale range
Linear range
±
150 % of full-scale range