5 interfaces, 1 rs 232 interface, 1 data transmission protocol – Metrohm 820 IC Separation Center User Manual
Page 87: Interfaces, Rs 232 interface, Data transmission protocol, N 5 interfaces
5.1 RS 232 interface
819 IC Detector / 820 IC Separation Center
79
5 Interfaces
5.1
RS 232 interface
The RS 232 interface 9 of the 819 IC Detector is used for the connection
of the 830 IC Interface or a PC (see Section 2.3.1 – 2.3.2) and allows
remote operation of 819 IC Detector and 820 IC Separation Center.
5.1.1
Data transmission protocol
The RS 232 interface 9 of the 819 IC Detector is configured as DTE
(Data Terminal Equipment) with the following technical specifications:
•
Standard
Data interface in accordance with EIA standard
RS 232C (DIN 66020, page 1)
•
Baud rate 9600
•
Data Bits
8
•
Stop Bits
1
•
XON/XOFF
on
•
Control characters C
R
DEC 13
HEX 0D
L
F
DEC 10
HEX 0A
XON
DEC 17
HEX 11
XOFF
DEC 19
HEX 13
•
Max. line length
80 characters + C
R
L
F
•
Mode
Full duplex (simultaneous transmission and re-
ceiving)
Restriction:
If data are first received by the interface, trans-
mission is not started until the receipt is at an
end.
•
Cable
For interconnection of the 819 IC Detector with
third-party devices, only a shielded data cable
may be used (e.g. Metrohm D.104.0201). The
cable shield must be perfectly earthed at both
devices (pay attention to current loops: always
use star earthing). Use only connectors with
sufficient shielding (e.g. Metrohm K.210.0001
with K.210.9004).
•
Cable length
max. ca. 15 m