4 Notes - Maintenance - Faults

819 IC Detector / 820 IC Separation Center

76

6 Insert connection piece

Insert connection piece 73 in suppressor holder 75 in such a

way that connection "1" is at the top, and that the three lugs

on the connection piece fit in the corresponding openings of

the holder.

Screw nut 72 onto the thread of suppressor holder 75 manu-

ally (do not use tools).

7 Connect and condition the suppressor

Reconnect the suppressor to the IC system.

Before switching the suppressor to the next position for the

first time, rinse all 3 suppressor units with solution for 5 min.