2.2 Installation of accessories
819 IC Detector / 820 IC Separation Center
17
•
Replace rear panel 21 and screw to the 820 IC Separation
Center using the four knurled screws 15.
2 Connect detector blocks
•
Plug the gray connecting cable permanently attached to the
detector block A into connection 6 „Detector Block“ of the
first 819 IC Detector and fasten to the instrument by
tightening the screws in the cable connector (see Fig. 9).
•
Plug the gray connecting cable permanently attached to the
detector block B into connection 6 „Detector Block“ of the
second 819 IC Detector and fasten to the instrument by
tightening the screws in the cable connector (see Fig. 9).
3 Connect waste container
•
Lead the outlet capillary of both detector blocks to a suffi-
ciently large waste container and fix in place.
Fig. 9:
Connection 2 x 819 Detector – 2.820.0X20
819
820
Cable from
detector block A
819
Cable from