3.1 819 IC Detector
819 IC Detector / 820 IC Separation Center
57
Program
On the
Program
subpage, program steps including time, program in-
struction and parameter can be entered.
First column
Time at which program instruction is applied.
Entry range:
0.0 ... 999.9 min
If no time is entered, the program instruction is
applied together with the last instruction with
time entry.
Second column
Program instruction (see below).
In addition to these predefined instructions,
user-defined remote commands can be en-
tered if activated on the
Remote configuration
tab.
Third column
Parameter for program instruction (see below).
ENABLED
Enable program start (a disabled program is
not started).
Add new program instruction.
Delete selected program instruction.
Test the time program (error messages are dis-
played if program is wrong).
List of program instructions
The following program instructions can be added to the time program
on the
Program
subpage:
Instruction Parameter entry
Meaning
Remote
0, 1, *
Set remote output lines 1...8 to the
desired values. For entry of the first
value, enter
1
,
0
or
*
. For entry of the
other values, move the cursor in front
of the value to be changed and enter
1
,
0
or
*
.
Range
100, 200, 5000
µ
S/cm
1, 2, 5, 10 mS/cm
Set measuring range to the selected
value.
FS 0.05
µ
S/cm ...
10 mS/cm
Set full scale range to the selected
value (depends on Range setting).
Zero on,
off
Switch on or off the auto-zero func-
tion.
Polarity +,
–
Reset the polarity of the output sig-
nal.
Mark
Trigger marking signal.