Fig. 8, Connection detector 819 – 2.820.0x10/2.820.0x30 – Metrohm 820 IC Separation Center User Manual
Page 24
2 Installation
819 IC Detector / 820 IC Separation Center
16
Fig. 8:
Connection Detector 819 – 2.820.0X10/2.820.0X30
2 Connect detector block
•
Plug the gray connecting cable permanently attached to the
detector block into connection 6 „Detector Block“ of the 819
IC Detector and fasten to the instrument by tightening the
screws in the cable connector (see Fig. 8).
3 Connect waste container
•
Lead the outlet capillary of the detector block to a sufficiently
large waste container and fix in place.
2.820.0X20 IC Separation Center
The two-channel versions 2.820.0X20 of the IC Separation Center can
be operated with two 819 IC Detectors. It is best to proceed as follows
when connecting the instruments and the two detector blocks:
1 Install detector blocks
•
Unscrew the four knurled screws 15 from the top rear panel
21 of the 820 IC Separation Center and remove rear panel
•
Position first detector block from the back in the space pro-
vided in the 820 IC Separation Center on the right and push
fully to the front (see Fig. 7, resp. Fig. 22 and Fig. 25).
•
Position second detector block from the back in the space
provided in the 820 IC Separation Center on the left and push
fully to the front (see Fig. 7, resp. Fig. 22 and Fig. 25).
•
Insert the cable permanently attached to the detector block A
in opening 20 and the outlet capillary in opening 19 "Waste A"
•
Insert the cable permanently attached to the detector block B
in opening 16 and the outlet capillary in opening 17 "Waste B"
819
820
Cable from
detector block