Fig. 26, Operating principle of auto-zero – Metrohm 820 IC Separation Center User Manual
Page 60
3 Operation
819 IC Detector / 820 IC Separation Center
52
Zero
Triggers the auto-zero function.
Zero
Switch off the auto-zero function.
"Auto-zero" is the name used for the automatic electronic back-
ground compensation, i.e. the current measured
Conductivity
value
is set to zero and hence lies in the middle of the selected full-scale
range.
The auto-zero function is disabled, if the actual value of the absolute
conductivity
abs. Cond.
out of the limit set by the measuring
Range
(see Fig. 27), the diagram below shows the operating principle of the
auto-zero function.
Fig. 26:
Operating principle of Auto-zero
Before the auto-zero function is activated for the first time or after it
has been switched off, the measuring signal lies outside the full-
scale range; it is automatically set to 0
µ
S/cm each time the
Zero
key is pressed.
The auto-zero function can be triggered at any time using the
Zero
key.
The auto-zero function works properly only when the measured
value remains relatively stable (e.g. not during the appearance of
the injection peak at the start of the chromatogram).
Zeit
0
+
Full Scale
–
Full Scale
+ 1.8 Full Scale
– 1.8
Full Scale
Conductivity
Conductivity
Out of range
Conductivity
Out of range
Conductivity
Out of range
Zero
Zero
Zero