Metrohm 820 IC Separation Center User Manual
Page 71
3.2 820 IC Separation Center
819 IC Detector / 820 IC Separation Center
63
Time program
The
Program
tab is only available for a
820.0XX0 Separation Center
con-
nected to a 830 IC Interface. The
820 IC Separation Center
that is con-
nected to a 819 IC Detector is controlled by the time program of the
detector.
On the
Program
tab of the
820.0XX0 Separation Center
window a user-
defined time program can be entered. This program is started auto-
matically as defined in the
Start mode
window (see Section 4.4.3 of the
«IC Net» Instructions for Use) either at the moment the determination is
started (
Start with determination
) or at the moment the sample is injected
(
Start with inject
).
First column
Time at which program instruction is applied.
Entry range:
0.0 ... 999.9 min
If no time is entered, the program instruction is
applied together with the last instruction with
time entry.
Second column
Program instruction (see below).
Third column
Parameter for program instruction (see below).
ENABLED
Enable program start (a disabled program is
not started).
Add new program instruction.
Delete selected program instruction.
Test the time program (error messages are dis-
played if program is wrong).