2 connection of column heating, Connection of column heating, Fig. 10 – Metrohm 820 IC Separation Center User Manual
Page 26: Install column heating in separation center, Ction 2.2.2 an
2 Installation
819 IC Detector / 820 IC Separation Center
18
2.2.2
Connection of column heating
The instrument versions 2.820.02X0 of the IC Separation Center contain
a column heating for thermostatting the separating column as an op-
tion. If you wish to use this column heating then it should be installed as
follows:
Fig. 10:
Install column heating in Separation Center
1 Laying the connection cable
•
Screw off the 4 knurled screws 15 from upper rear panel 21
of the 820 IC Separation Center and remove the rear panel
•
Lead 6.2108.120 Cable through opening 20.
•
Replace rear panel 21 and screw it onto 820 IC Separation
Center using the 4 knurled screws 15.
2 Insert column heating
•
Carefully place the column heating inside the Separation
Center. If the column has already been inserted in the heating
then be careful with the capillaries at feedthroughs 35 and 41
(see Section 2.9.2).
•
Use magnetic holder 39 and the corresponding counter-
pieces to attach the column heating to the left-hand inside
wall of the Separation Center.
3 Connect column heating
•
Insert 6.2108.120 Cable at connection 38 of the column heat-
ing (see Fig. 6) and at connection 23 on the rear panel of the
Separation Center (see Fig. 5).