Fig. 27, Diagram of measuring and full-scale range – Metrohm 820 IC Separation Center User Manual
Page 62
3 Operation
819 IC Detector / 820 IC Separation Center
54
Full
scale
The full scale range (operating range) sets the
desired sensitivity for the analog output. The
possible values of the full scale range in
µ
S/cm
depend on the preset measuring
Range
.
This value can be reduced in a total of 11 stages by a factor 1 (low-
est sensitivity) up to maximum a factor of 2000 (highest sensitivity).
The full-scale range is primarily used to increase the sensitivity for
the recording of the chromatogram following the electronic back-
ground compensation of the eluent sensitivity (auto-zero function,
see also description of the auto-zero function above.
For the selected full-scale range (FS), the linearity of the conductivity
measurement is assured in the range –1.8 Full Scale
…
+1.8 Full
Scale. If the measured value violates these limits, the live display of
Conductivity
in the
WATCH WINDOW
changes it’s color to the settings
for
Out of range
. (default: red). At the same time the absolute conduc-
tivity is displayed instead of the auto-zero value.
The full-scale range determines the limits for the output of the analog
…
1 V). The polarity of the output signal
can be changed at any time with the parameter
Polarity
to
+
(0
…
+Full Scale) or
–
(0
…
–Full Scale).
Fig. 27:
Diagram of measuring and full-scale range
Full-scale-range
Measuring range
Analog output
Conductivity
0
R
1.1 R
0
+ FS
– FS
+ 1.8 FS
– 1.8 FS
0
+
–
0 V
10 mV
10 mV
Conductivity Out of Range
abs. Cond.
Out of Range
Conductivity Out of Range