Care and use manual – Waters ACQUITY UPLC BEH Columns User Manual
Page 9
4. Avoid (where possible):
■
■
100% aqueous mobile phases (if possible).
■
■
HPLC-grade bottled water.
■
■
“Topping off” or adding “new” mobile phase to “old” mobile phase.
■
■
Old aqueous mobile phases. Remember to rinse bottles thoroughly and prepare fresh every 24 to 48 hours.
■
■
Using phosphate salt buffer in combination with high ACN concentrations (e.g., > 70%) due to precipitation.
5. Don’t: assume a “bad” column is the culprit when high backpressure or split peaks are observed:
Investigate cause of column failure
■
■
Backpressure
■
■
Mobile phase(s), bacteria, precipitation and/or samples
■
■
Peak splitting
■
■
Sample quality
■
■
Injection solvent strength.
6. Remember: the diameter of UPLC columns (1.0, 2.1 and 3.0 mm ID) are often lower than that of a conventional HPLC column and
therefore, mobile phases last much longer. To reduce the chances of mobile-phase contamination or degradation, only prepare what you
need for analysis or store excess bulk quantities in a refrigerated environment.
7. Mobile phase-related questions to ask:
■
■
Am I using 100% aqueous mobile phases? Am I able to add a small amount of organic modifier to my mobile phase A?
■
■
Do I filter my aqueous mobile phases through 0.2 µm filters?
■
■
How old is my mobile phase? Do I label the bottle with preparation date?
■
■
Do I “top off” or do I prepare fresh mobile phases every 24 – 48 hours?
■
■
What is the quality of my water? Has the quality recently changed? How is my water purification system working? When was it
last serviced?
■
■
Am I working with pH 7 phosphate buffer (which is VERY susceptible to bacterial growth)?
8. Sample-related questions to ask:
■
■
If I inject neat standards prepared in mobile phase do I observe these problems?
■
■
If I prepare my standards in water and prepare them like samples (e.g., SPE, filtration, etc.) do I still observe these problems?
■
■
Has the quality of my samples changed over time?
b. Rcommended Flow Rates and Backpressures for Reversed-Phase ACQUITY UPLC BEH Columns
1.0 mm i.d. Columns (40 °C)
UPLC Linear Velocity
(mm/sec)
3
4
5
6
Column Dimensions
Flow Rate
(mL/min)
Backpressure
(psi)
Flow Rate
(mL/min)
Backpressure
(psi)
Flow Rate
(mL/min)
Backpressure
(psi)
Flow Rate
(mL/min)
Backpressure
(psi)
1.0 x 50 mm
0.1
4300
0.13
5600
0.17
7400
0.2
8700
1.0 x 100 mm
0.1
8600
0.13
11200
0.17
14600
0.2
17200
1.0 x 150 mm
0.1
12800
0.13
16700
0.17
21800
0.2
25600
9
[ CARE AND USE MANUAL ]