Care and use manual – Waters ACQUITY UPLC BEH Columns User Manual

Page 9

background image

4. Avoid (where possible):

100% aqueous mobile phases (if possible).

HPLC-grade bottled water.

“Topping off” or adding “new” mobile phase to “old” mobile phase.

Old aqueous mobile phases. Remember to rinse bottles thoroughly and prepare fresh every 24 to 48 hours.

Using phosphate salt buffer in combination with high ACN concentrations (e.g., > 70%) due to precipitation.

5. Don’t: assume a “bad” column is the culprit when high backpressure or split peaks are observed:

Investigate cause of column failure

Backpressure

Mobile phase(s), bacteria, precipitation and/or samples

Peak splitting

Sample quality

Injection solvent strength.

6. Remember: the diameter of UPLC columns (1.0, 2.1 and 3.0 mm ID) are often lower than that of a conventional HPLC column and

therefore, mobile phases last much longer. To reduce the chances of mobile-phase contamination or degradation, only prepare what you
need for analysis or store excess bulk quantities in a refrigerated environment.

7. Mobile phase-related questions to ask:

Am I using 100% aqueous mobile phases? Am I able to add a small amount of organic modifier to my mobile phase A?

Do I filter my aqueous mobile phases through 0.2 µm filters?

How old is my mobile phase? Do I label the bottle with preparation date?

Do I “top off” or do I prepare fresh mobile phases every 24 – 48 hours?

What is the quality of my water? Has the quality recently changed? How is my water purification system working? When was it
last serviced?

Am I working with pH 7 phosphate buffer (which is VERY susceptible to bacterial growth)?

8. Sample-related questions to ask:

If I inject neat standards prepared in mobile phase do I observe these problems?

If I prepare my standards in water and prepare them like samples (e.g., SPE, filtration, etc.) do I still observe these problems?

Has the quality of my samples changed over time?

b. Rcommended Flow Rates and Backpressures for Reversed-Phase ACQUITY UPLC BEH Columns

1.0 mm i.d. Columns (40 °C)

UPLC Linear Velocity

(mm/sec)

3

4

5

6

Column Dimensions

Flow Rate

(mL/min)

Backpressure

(psi)

Flow Rate

(mL/min)

Backpressure

(psi)

Flow Rate

(mL/min)

Backpressure

(psi)

Flow Rate

(mL/min)

Backpressure

(psi)

1.0 x 50 mm

0.1

4300

0.13

5600

0.17

7400

0.2

8700

1.0 x 100 mm

0.1

8600

0.13

11200

0.17

14600

0.2

17200

1.0 x 150 mm

0.1

12800

0.13

16700

0.17

21800

0.2

25600

9

[ CARE AND USE MANUAL ]