7. If acetone is used in one or more mobile phases, do not use

acetone as a sample diluent or needle wash solvent. Refer to
injection solvents section for sample diluent recommendations
and miscellaneous tip (#3) for needle wash solvent/purge
solvent recommendations.

8. Typical sample preparation suggestions for samples that contain

sugars/saccharides/carbohydrates:

a. Liquid Samples

i. Dilute with 50:50 ACN/H

2

O

ii. Filter using 0.45 µm or 0.22 µm syringe filter

(if necessary)

b. b. Solid Samples

i. Weigh out sample (~3 g) into 50 mL centrifuge tube

ii. Add 25 mL of 50:50 ACN/H

2

O and homogenize

(mechanically)

iii. Centrifuge at 3200 rpm for 30 minutes

iv. Collect supernatant and filter using 0.45 µm or

0.22 µm syringe filter (if necessary)

c. Depending on sample and/or analyte concentrations,

additional sample dilutions may be necessary.

d. More complex samples and/or lower analyte

concentrations may require additional sample
preparation steps and/or procedures such as solid phase
extraction (SPE).

e. Consider VanGuard BEH Amide pre-columns for UPLC

column protection.

Waters Corporation
34 Maple Street
Milford, MA 01757 U.S.A.
T: 1 508 478 2000
F: 1 508 872 1990
www.waters.com

[ CARE AND USE MANUAL ]

Waters, The Science of What’s Possible, ACQUITY UPLC, Oasis, Sep-Pak, Empower, MassLynx, Atlantis, and UPLC are registered
tradmarks of Waters Corporation. eCord and VanGuard are trademarks of Waters Corporation. All other trademarks are the property of
their respective owners.

©2014 Waters Corporation. Produced in the U.S.A. November 2014 Rev H 715001371 KP-PDF