BioSuite Columns

4

IV. Waters BIosuIte Phenyl hIc columns

This section contains recommended operating conditions and specifica-
tions for the Waters polymer-based, Phenyl Hydrophobic-interaction
Chromatography columns.

Operating Conditions

1. Shipping Solvent: HPLC grade Water

2. Recommended Operating Flow Rates/Max Pressure:

Note: When using a high viscosity buffer and ambient temperature, the maxi-
mum flow rate may have to be reduced to avoid exceeding the maximum
recommended operating pressure for the column. When changing solvents,
use 50% of the maximum flow rate (20% for the 21.5 mm i.d. preparative
column).

3. Recommended pH Range: 2 - 12 (Avoid extended operation at pH > 10

to maximize column life.)

4. Recommended Organic Conc.: less than 50% (Note: Reduced organic

concentrations may be required if buffers are present in the eluent in
order to avoid salt precipitation.)

5. Recommended Salt Conc.: Less than or equal to 4.0 Molar

6. Recommended Temperature: 10-50 °C. Reduce flow rate when

operating at low temperatures (e.g. 10 °C) to avoid excessive column
pressure.

7. Recommended Cleaning Solvents/Procedures:

A) 0.1 - 0.2M NaOH, and, if not successful,

B) 20 - 40% acetic acid in HPLC grade water.

Clean the column regularly by injecting up to one column volume 0.1
- 0.2M NaOH in 250 µl increments. Use proportionally larger volumes
for the 21.5 mm i.d. preparative column. Following use of cleaning
solvent, rinse the column with 3 to 5 column volumes of HPLC grade
water before storage.

8. Recommended Storage: For overnight storage, continuously flush the

column with the mobile phase at 10 - 20% of the maximum recom-

mended flow rate. Store the column in the HPLC grade water
when it will be used within 24 hrs or in either 20% ethanol or metha-
nol for long term storage. (Caution: Rinse column in HPLC

grade water before switching to 20% organic solvent to avoid salt
precipitation in column.)

V. Waters BIosuIte ultra-hIgh resolutIon (uhr), hIgh
resolutIon (hr), and sIze-exclusIon (sec) columns

This section contains recommended operating conditions and specifications
for the Waters silica-based, ultra-high Resolution (UHR), high-resolution (HR),
and standard size-exclusion (SEC) columns.

Operating Conditions

1. Shipping Solvent: 0.05% Sodium Azide or 0.1M Sodium Sulphate in

0.1M phosphate buffer, pH 6.7

2. Recommended Operating Flow Rate/Max Pressure:

Description

Matrix

i.d.

(mm)

L

(mm)

Flow Rates /

Max Pressure

BioSuite Phenyl, 1000,
10 µm HIC

polymer

7.5

75

0.5 - 1.2 mL/min /

300 PSI

BioSuite Phenyl, 1000,
13 µm HIC

polymer 21.5

150

4.0 - 8.0 mL/min /

300 PSI

Description

Matrix

i.d.

(mm)

L

(mm)

Flow Rates /

Max Pressure

BioSuite 125, 4 µm
UHR SEC

silica

4.6

300

0.1 - 0.4 mL/min /

1800 PSI

BioSuite 250, 4 µm
UHR SEC

silica

4.6

300

0.1 - 0.4 mL/min /

1800 PSI

BioSuite UHR Guard SEC

silica

4.6

35

0.1 - 0.4 mL/min /

1800 PSI

BioSuite 125, 5 µm
HR SEC

silica

7.8

300

0.5 - 1.2 mL/min /

1050 PSI

BioSuite 250, 5 µm
HR SEC

silica

7.8

300

0.5 - 1.2 mL/min /

1050 PSI

BioSuite 450, 8 µm
HR SEC

silica

7.8

300

0.5 - 1.2 mL/min /

525 PSI

BioSuite HR Guard SEC

silica

6

40

0.5 - 1.2 mL/min /

1050 PSI

BioSuite 125, 10 µm
SEC

silica

7.5

300

0.5 - 1.2 mL/min /

300 PSI

BioSuite 125, 13 µm
SEC

silica

21.5

300

3.0 - 8.0 mL/min /

150 PSI

BioSuite 250, 10 µm
SEC

silica

7.5

300

0.5 - 1.2 mL/min /

375 PSI