beautypg.com

Care and use manual, Xi. warrant y, Xii. ordering information – Waters Dextro-Pak Cartridge User Manual

Page 7

background image

[ Care and Use ManUal ]

Dextro-Pak Cartridge

7

Table Troubleshooting Procedures

XI. warrant y

Waters warrants its high-performance liquid chromatography
cartridges in accordance with the following terms and conditions:
Waters will repack or replace (at our discretion) without cost, any
cartridge that fails to perform satisfactorily if notified within 60 days
from receipt of the cartridge. Any cartridge returned must have a
prior Return Authorization Number assigned by the Waters Customer
Service Department. Approval is subject to the following exclusions:

Physical damage to the cartridge because of misuse or abuse.

Chemical damage to the packing material because of use with
incompatible solvents or buffers, or at an incorrect pH.

• Physical damage to the packing material because of opera-

tion at incorrect temperatures or pressures.

High internal pressures which have occurred in the cartridge
due to the improper solvent or sample filtration practices
causing particulate buildup or precipitation in the cartridge or
end fittings due to high internal pressures which have occurred
because of improper solvent or sample filtration practices.

XII. orderIng InformatIon

Problem

Cause

Solution

Excess pressure buildup

In-line filters lugged with particulates

Replace the in-line filter or shut off the pump. When inlet pressure has
dropped to zero, disconnect the in-line filter from the cartridge and
backflush the prefilter only with mobile phase at 9 ml/min until the
particulates are backflushed out.

Excess pressure buildup

Guard column clogged with particles or adsorbed materials

Replace spent guard column with fresh packing.

Fluctuating backpressure

Cartridge inlet frit plugged with particulates.

Always use an in-line filter or guard column to prevent this. Reverse
cartridge to try to wash out particulates at normal flow rate.

Gas in mobile phase

Check degassing procedure.

“Spurious” peaks, not due to sugars

Faulty pump operation Elution of salts and/or acids

Keep mobile phase heated to 70 °C. Inspect pump check valves and
carry out “Ramp Test” (consult the operator’s manual for your pump).
Check sample cleanup procedures.

Leaking mobile phase

Fittings in bad state

Tighten fittings properly.

Variable elution times of repair

Variations in flow rate

DO NOT OVERTIGHTEN. Replace worn fittings,o-rings, or seals. Check
system for faults, leaks, especially pulsations from pump.

Band broadening or loss of resolution

Guard column fouled

Replace spent guard column with fresh packing.

Analytical cartridge fouled or old

Try washing with 20% acetonitrile, 80% water. If cartridge does not
recover, use a new cartridge.

Item

Part Number

Compression screws and ferrules (5 pk)

WAT025604

Guard-Pak~” Inserts Resolve”’ C1e (10 pk)

WAT085824