Styragel Columns

9

2) Column Efficiency Test

The column-efficiency test described below may be used to calculate a theoreti-
cal plate count, which is a measurement of column efficiency within your system.

Solvents

It is not necessary to use the solvent that your column is shipped in to deter-
mine the efficiency of your column. Initial plate counts that are determined
using different solvents will have different values. Test columns individually
using your normal operating solvent.

Test procedure

To perform a column efficiency test:

1. Slowly increase the flow rate to 1.0 mL/min for the HR 0.5,

HR 1, HR 2, and HR 4E columns, or to 2.0 mL/min for all
other columns.

2. Adjust the detector to an attenuation that achieves a peak of 70

percent full scale (noise level <0.5 percent full scale).

3. Set the recorder chart speed to 50 mm/min, or the data system

sampling rate to at least 5 data points per second.

4. Inject up to 20 µL of marker solution. Use a solution of up to 10

percent marker in solvent.

Use up to 5 µL for solvent efficient

columns.

Refer to Table 4, Table 5, or Table 6 for the marker to use when performing
the efficiency test.

5. Record the retention time, instrument settings, and column con-

figuration so you can reproduce them exactly for comparison in
the future.

6. Compute the plate count using the tangent method (see

Figure 6). Use these results for comparison throughout the life
of your column.

7. Replace any column that exhibits a plate count more than 30

percent below the original value.

Table 4. Conditions and sample for 7.8 mm for Column Efficiency Test:
HR Columns

1

Dicyclohexylphthalate

Table 5. Conditions and Sample for 4.6 mm Column Efficiency Test:
HR Columns

Column

Type

Toluene

THF

DMF

Flow Rate

mL/min

HR 0.5

Acetone

Propylbenzene

Ethyleneglycol

1.0

HR 1

Acetone

Propylbenzene

Ethyleneglycol

1.0

HR 2

Acetone

Propylbenzene

Ethyleneglycol

1.0

HR 3

ODCB

Acetone

Acetone

2.0

HR 4

ODCB

Acetone

Acetone

2.0

HR 4E

Acetone

Propylbenzene

Ethyleneglycol

1.0

HR 5E

DCHP2

Acetone

Acetone

2.0

Column

Type

Toluene

THF

DMF

Marker

Flow Rate

(mL/min)

Marker

Flow Rate

(mL/min)

Marker

Flow Rate

(mL/min)

HR 0.5

ODCB

0.3

Acetone

0.3

Acetone

0.3

HR 1

ODCB

0.3

Acetone

0.3

Acetone

0.3

HR 2

ODCB

0.3

Acetone

0.3

Acetone

0.3

HR 3

ODCB

0.3

Acetone

0.3

Acetone

0.3

HR 4

ODCB

0.3

Acetone

0.3

Acetone

0.3

HR 4E

ODCB

0.3

Acetone

0.3

Acetone

0.3

HR 5E

DCHP

2

0.3

Acetone

0.3

Acetone

0.3