[ Care and Use ManUal ]

Styragel Columns

7

v. care and maIntenance

This chapter describes:

• Troubleshooting

• Storing the column

• Efficiency testing

a. Troubleshooting

Table 3 describes specific problems, causes, and corrective actions.

Styragel

®

HT

Molecular W

eight

Styragel

®

HT 2

Elution Volume (mL)

Styragel

®

HT 3

Styragel

®

HT 4

Styragel

®

HT 5

Styragel

®

HT 6

Styragel

®

HT 6E

Styragel

®

HT

Molecular W

eight

Styragel

®

HT 2

Elution Volume (mL)

Styragel

®

HT 3

Styragel

®

HT 4

Styragel

®

HT 5

Styragel

®

HT 6

Styragel

®

HT 6E

Figure 4. Calibration Curves for Waters Styragel HMW Columns

Styragel

®

HMW

Molecular W

eight

Styragel

®

HMW 2

Elution Volume (mL)

Styragel

®

HMW 6E

Styragel

®

HMW 7

Styragel

®

HMW

Molecular W

eight

Styragel

®

HMW 2

Elution Volume (mL)

Styragel

®

HMW 6E

Styragel

®

HMW 7

Problem

Cause

Corrective Action

Buildup in system operating

pressure

Inlet filter insert (from the first column only) plugged with par-

ticulates Injector and pump seal shedding

Replace the inlet filter insert.

Install a Styragel Guard column.

Fit an in-line filter between the pump and the

first column.

Clogged tubing

Replace the tubing

Loss of resolution, broad peaks,

low plate counts

Filter inserts partially blocked

Replace or clean inlet

and/or outlet filter inserts

Failing injector

Repair the injector

Insufficient equilibrium

Continue equilibrium

Column damaged

Replace column

Table 3. Column Troubleshooting

Figure 3. Calibration Curves for Waters Styragel HT Columns