Care and use manual, V. care and maintenance, A. troubleshooting – Waters Styragel Columns User Manual
Page 7

[ Care and Use ManUal ]
Styragel Columns
7
v. care and maIntenance
This chapter describes:
• Troubleshooting
• Storing the column
• Efficiency testing
a. Troubleshooting
Table 3 describes specific problems, causes, and corrective actions.
Styragel
®
HT
Molecular W
eight
Styragel
®
HT 2
Elution Volume (mL)
Styragel
®
HT 3
Styragel
®
HT 4
Styragel
®
HT 5
Styragel
®
HT 6
Styragel
®
HT 6E
Styragel
®
HT
Molecular W
eight
Styragel
®
HT 2
Elution Volume (mL)
Styragel
®
HT 3
Styragel
®
HT 4
Styragel
®
HT 5
Styragel
®
HT 6
Styragel
®
HT 6E
Figure 4. Calibration Curves for Waters Styragel HMW Columns
Styragel
®
HMW
Molecular W
eight
Styragel
®
HMW 2
Elution Volume (mL)
Styragel
®
HMW 6E
Styragel
®
HMW 7
Styragel
®
HMW
Molecular W
eight
Styragel
®
HMW 2
Elution Volume (mL)
Styragel
®
HMW 6E
Styragel
®
HMW 7
Problem
Cause
Corrective Action
Buildup in system operating
pressure
Inlet filter insert (from the first column only) plugged with par-
ticulates Injector and pump seal shedding
Replace the inlet filter insert.
Install a Styragel Guard column.
Fit an in-line filter between the pump and the
first column.
Clogged tubing
Replace the tubing
Loss of resolution, broad peaks,
low plate counts
Filter inserts partially blocked
Replace or clean inlet
and/or outlet filter inserts
Failing injector
Repair the injector
Insufficient equilibrium
Continue equilibrium
Column damaged
Replace column
Table 3. Column Troubleshooting
Figure 3. Calibration Curves for Waters Styragel HT Columns