Care and use manual – Waters Sugar-Pak I Column User Manual
Page 3

[ Care and Use ManUal ]
Sugar-Pak I Columns
3
foil to minimize evaporation. Bring the mobile phase to 
its boiling point for a few minutes just before use, but 
maintain the temperature between 70-90 °C during use. 
This practice ensures that gases (especially C0
2
) do not
redissolve in the mobile phase. It also prevents the growth of 
microorganisms. 
Keep the mobile phase reservoir clean and covered and supply
freshly prepared mobile phase every 24 hours.
c. Equilibration
To ensure that a new column is fully equilibrated with calcium prior to 
analysis, the following procedure is recommended.
1. Install the column with the normal direction of flow reversed
 (arrow on label pointing to column inlet tubing). A backflush 
 valve may also be used.
2. Back flush the column with at least 100 mL of 0.001 M calcium
 EDTA solution at 90 °C using a flow rate of 0.5 mL/min. A 
 concentration of 500 mg/L of calcium EDTA approximates 
 0.001 M.
3. Reverse the flow direction again (back to the normal direction)
 and flush the column with the stable baseline indicating that the 
 column is ready for use.
III. CARe AnD Use
Liquid chromatography columns have a finite life influenced by their 
care and use, number of injections, sample and solvent cleanliness, 
frequency of solvent changeover, and handling and storage procedures 
(among other factors). If a change is observed in the:
retention of a particular compound,
resolution between two compounds, or
peak shape
take immediate steps to determine the reason for the changes. Until you 
have made this determination, you must not rely upon the results of any 
separation using the column. Follow generally accepted procedures for 
quality control and methods development when using these columns. 
Note: Before running the first analysis on your new column, perform 
the test sample separation given in the test conditions section.
Before using your new Sugar-Pak I column for the first time, it is 
recommended that the following procedure be observed:
1. Connect the column (refer to Section II a)
2. Set the flow rate at 0.2 mL/min until the column temperature
reaches 70 °C.
3. Once the column temperature reaches 70 °C, increase the
 flow rate to 0.6 mL/min in steps of 0.1 mL/min. Wait for the 
 backpressure to stabilize between each 0.1 mL/min increase.
a. Precautions
Normal recommended pressure should not exceed: 2000 psi.
Maximum flow rate: 0.6 mL/min (at or above 70 °C).
DO NOT use calcium chloride, calcium nitrate, or any other acidic
calcium salt in the mobile phase, as this can corrode the column 
and damage the packing.
Maximum mobile-phase organic content is 5% V/V. Small amounts
of acetonitrile, ethanol, methanol, and isopropanol in the sample 
will not effect column performance.
Column temperatures should not exceed 95 °C. Generally, high
temperatures provide greater resolution, but little improvement 
occurs between 90 °C and 95 °C. Temperatures below 70 °C do not 
provide adequate resolution of many sugars due to the separation 
of anomers. For quantification of ethanol, a column temperature of 
75 °C should be used.
Column temperature changes can be rapid without adverse
effects. If the column temperature is below 60 °C, a maximum 
flow rate of 0.2 mL/min should be used. Higher flow rates may 
produce excessive backpressure due to the higher viscosity of 
water at lower temperatures.
Reverse flow direction through the column (after every 5 liters
of mobile phase). This procedure will prolong column life 
significantly. Flow rate changes should be made slowly and not 
exceed 0.5 mL/min
2
(0.5 mL/min per minute). It is advisable to
change the flow rate in steps of 0.1 mL/min and wait until the 
backpressure stabilizes before the next step change. 
