Care and use manual – Waters XBridge Protein BEH, C4, 300A, 3.5 µm Columns User Manual

Page 3

background image

3

[ CARE AND USE MANUAL ]

XBridge Protein BEH C

4

, 300Å, 3.5, 5, 10 �m

d. Useful Functional Tests for Benchmarking a New Column

The Column Efficiency Test described above is a useful measure
of the physical state of the packed bed as well as an indicator
of the chemical integrity of the bonded phase. It may also be
useful to benchmark the column performance with a sample that is
more representative of the intended application. Two tests can be
suggested as starting points for benchmarking a new column and
for monitoring a column during its use.

Peptide Mixture Performance Test

Sample:

Waters MassPREP™ Peptide Standard Mixture

(P/N 186002337)

Reconstitute 1 vial in 100 μL 0.1% TFA:5%

Acetronitrile: 94.9% Water

Injection Volume: 2.1 mm – 3.3 µL

4.6 mm – 16.0 µL

Column:

XBridge Protein BEH C

4

, 300Å

3.5 µm, 2.1 x 50 mm

Flow Rate:

2.1 mm – 0.2 mL/min

4.6 mm – 0.96 mL/min

Mobile Phase:

A: 0.1% TFA in water

B: 0.075% in 71.4% acetonitrile/

28.6% water

Gradient:

Gradient Time for Different Column Lengths

50 mm 100 mm 150 mm 250 mm

%A

%B

Curve

Initial

Initial

Initial

Initial

100

0

*

30

60

90

150

30

70

6

32

64

96

160

30

70

1

50

100

150

250

100

0

1

Temperature:

40 °C

Detection:

220 nm

Figure 1: Typical Peptide Chromatogram Using MassPREP Peptide Standard Mixture

This chromatogram is typical of the results obtained in Waters
laboratories with the method described above, using a XBridge
Protein BEH C

4

, 300Å, 3.5 µm, 2.1 x 50 mm Column. The

retention times will double, triple, and be five times greater
for the 100 mm, 150 mm and 250 mm columns respectively.
The exact results observed in any laboratory will depend on
the instrument in use. System volume, gradient generation
mechanism, mixing, design of temperature control, detector cell
dimensions, detector optical properties, and detector electronic
properties all have a direct impact on the observed chromatogram.
The pattern should be similar, however, on any well-functioning,
modern HPLC. This test is exceptionally valuable for monitoring
the life of the column and for troubleshooting separation
difficulties that may arise.

Protein Mixture Performance Test

Sample: MassPREP Protein Standard Mixture (P/N 186004900)

Dissolved in 0.1% TFA:5% acetonitrile:94.9% water

Protein

Sigma P/N

Conc. mg/mL

Bovine Ribonuclease A

R5500

0.04

Horse Cytochrome c

C7752

0.06

Bovine Serum Albumin

A8022

0.20

Horse Myoglobin

M1882

0.13

Yeast Enolase

E6126

0.22

Rabbit Phosphorylase b

P6635

0.59

Injection Volume: 2.1 mm – 5.0 µL

4.6mm – 25.0 µL

Column:

XBridge Protein BEH C

4

, 300Å

,

3.5 µm, 2.1 x 50 mm

Flow Rate:

2.1 mm – 0.2 mL/min

4.6 mm – 0.96 mL/min