[ CARE AND USE MANUAL ]

XBridge Protein BEH C

4

, 300Å, 3.5, 5, 10 �m

use a Keystone, Inc. SLIPFREE

®

connector to always ensure the

correct fit. The finger tight SLIPFREE connectors automatically
adjust to fit all compression screw type fittings without the use of
tools (Figure 7).

Figure 7: Single and Double SLIPFREE Connectors

SLIPFREE Connector Features:

§

Tubing pushed into endfitting, thereby guaranteeing a

void-free connection

§

Connector(s) come(s) installed on tubing

§

Various tubing i.d.s and lengths available

§

Fingertight to 10,000 psi – never needs wrenches

§

Readjusts to all column endfittings

§

Compatible with all commercially available endfittings

§

Unique design separates tube-holding function from

sealing function

Table 3: Waters Part Numbers for SLIPFREE Connectors

SLIPFREE Type

Tubing Internal Diameter

Tubing
Length

0.005”

0.010”

0.020”

Single 6 cm

PSL 618000

PSL 618006

PSL 618012

Single 10 cm

PSL 618002

PSL 618008

PSL 618014

Single 20 cm

PSL 618004

PSL 618010

PSL 618016

Double 6 cm

PSL 618001

PSL 618007

PSL 618013

Double 10 cm

PSL 618003

PSL 618009

PSL 618015

Double 20 cm

PSL 618005

PSL 618001

PSL 618017

Minimizing Band Spreading

Figure 8 shows the influence of tubing internal diameter on
system band spreading and peak shape. As can be seen, the
larger tubing diameter causes excessive peak broadening and
lower sensitivity.

Diluted/Distorted Sample Band

0.005 inches

0.020 inches

0.040 inches

Figure 8: Effect of Connecting Tubing on System

b. Measuring System Variance and System Volume

This test should be performed on an HPLC system with a UV detector.

1. Disconnect column from system and replace with a zero dead

volume union.

2. Set flow rate to 1 mL/min.

3. Dilute a test mix in mobile phase to give a detector sensitivity

of 0.5–1.0 AUFS (system start up test mix can be used which
contains uracil, ethyl, and propyl parabens; Waters Part
Number WAT034544).

4. Inject 2 to 5 μL of this solution.

5. Measure the peak width at 4.4% of peak height

(5-sigma method):

5-sigma Bandspreading (μL) = Peak Width (min) x Flow

Rate (mL/min) x (1000 μL/1 mL)

System Variance (μL

2

) = (5-sigma bandspreading)

2

/ 25

System Volume

4.4% h

Figure 9: Determination of System Bandspreading Volume Using 5-Sigma Method